Instrument Settings

Computing Instrument Settings from Instrument Specifications

This article will explain the most commonly used methods for computing instrument standard errors, or "Instrument Settings" for STAR*NET. An online calculator that applies these formulas is available here.

Important Points:

  • Direction Setting weights direction sets (DB, DN, DE combinations) and DV
  • Direction Setting can be set a number of ways:
    • Determine by field testing
    • Use the Manufacturer’s DIN 18723 Specification for Direction Accuracy. See here for more information.
    • Use the Manufacturer’s ISO 17123-3 Specification for Direction Accuracy
  • Angle Setting weights M, SS, A measurements.
    • Angular settings can be determined from field testing
    • Angular settings can be calculated as shown below
  • Azimuth/Bearing weights B and BM measurements
  • Distance and constant settings weight D, DV or any distance component
  • Zenith and Vertical Difference settings are derived from direction setting and (sometimes) distance constant.
    • NOTE: Vertical Difference Settings only need to be entered when the "Distance/Vertical Data Type" option on the General tab of project settings is set to "Horiz Dist / Elev Diff"
  • Centering affects most measurements
Direction Accuracy:
DIN 18723 Specification for Theodolite Accuracy (i.e. a “1 second instrument”) applies to a repeated F1 and F2 “Pointing” or target sighting (ISO 17123 Specification is the same for practical purposes.)
Direction Std Err = DIN 18723 (or ISO 17123)
Or if direction is averaged from N observations::
Std Error = DIN 18723 / √ N
Eg: 2 observations with a 1 second instrument:
1”/ √ 2 = .71”
Eg: 6 observations with a 2 second instrument:
2”/ √ 6 = .82”
Angular Accuracy:
Angular Std Err = 2 x (DIN 18723) / √Number of “Pointings”
ie: 1 BS and one FS = 2 "pointings"
E.g.: 1 sets (1 FS + 1 BS) with a 1 second instrument: 2(1”) / √2 = 1.414”
E.g.: 3 sets (3 FS + 3 BS) with a 5 second instrument: 2(5”) / √6 = 4.082”
Distance Accuracy
Given as the Constant and PPM values generated when the EDM was calibrated. When multiple observations are averaged the constant must be calculated:
Where:
Constant = Given EDM Constant
PPM = Given ppm constant divided by 1 million
N = number of distance measurements averaged
Distance Constant = Constant/ √N
PPM is as given, it is not affected by averaging
Eg.: EDM with PPM = 5, constant = .005 meters averaging 2 measurements:
Constant = .005m / √2 = .00353m, PPM = 5
Eg: EDM with PPM = 1, constant = .002 meters averaging 15 measurements:
Constant = .002m /√15 = .000516 meters, PPM = 1
NOTES:
  1. If PPM and constant are only available in metric, switch your project units to Meters to enter these values, and then switch back to feet.
  2. PPM is the same regardless of number of observations.


Azimuth/Bearing
Directional accuracy to be applied to given directions, gyroscopic or astronomical observations.

Zenith
Directional accuracy of zenith angles. Typically 2 x the direction setting.
Where:
N = number of distance measurements averaged
Std Error = (2*DIN 18723) / √N
Eg: 2 observations with a 1 second instrument:
2*1”/ √ 2 = 1.414”
Eg: 6 observations with a 2 second instrument:
2*2”/ √ 6 = 1.633”

Elev Diff Constant and Elev Diff PPM
ONLY takes effect if you are manually entering 3D data in the form of horizontal distances and vertical differences. Not used very often.
Elev Diff Constant: use your judgement, it just has to be non-zero
Elev Diff PPM: if Elev Diff was computed from zenith observations and slope distances you can derive a PPM after computing the zenith value above.
PPM=sin(zenith)*1000000
Eg: 2 observations with a 1 second instrument:
Sin(1.414”)*1000000 = 6.856 PPM
Eg: 6 observations with a 2 second instrument:
Sin(1.633”)*1000000 = 7.95 PPM
Elev Diff PPM values applied to elev diff input compare well to zenith values applied to zenith/slope distance input for computed Z standard deviataions in the listing when slopes are shallow, but are less effective for steep observations.
NOTES:
  1. If PPM and constant are only available in metric, switch your project units to Meters to enter these values, and then switch back to feet.
Centering Errors
Horiz. Instrument: How well is the instrument centered over the mark?
Horiz. Target: How well is a target prism centered over the mark?
  • If using robotic aiming you may need to propogate ATR centering and tribrach centering together:
  • Horiz Target = √(ATR²+tribrach²)
Vertical: How well is the height of target or Instrument measured?